INPUT |
Voltage |
115/230 VAC ± 10%, User selection |
Frequency |
50/60 Hz ± 5% |
Fuse |
6.3 Amp 250V Slo-Blo |
DIELECTRIC WITHSTAND TEST MODE |
Output Rating |
5 KV @ 20mA AC |
6 KV @ 5 mA DC |
Output Adjustment |
Range: |
0-5 KV AC, 0-6 KV DC |
Resolution: |
10 volt/step |
Accuracy: |
± (2% of setting + 5 volts) |
Voltage Display |
Range: |
0 - 5 KVAC / 0-6KV DC |
Resolution: |
10 volt/step |
Accuracy: |
± (2% of reading + 10 volts) |
HI / LO -Limit AC |
Range: |
0.00 - 20.00 mA |
Resolution: |
0.01 mA/step |
Accuracy: |
± (2% of setting + 0.02 mA) |
AC Current Display |
Range: |
0.00 - 20.00 mA |
Resolution: |
0.01 mA/step |
Accuracy: |
± (2% of reading + 0.02 mA) |
HI / LO -LimitDC |
Range: |
0.00 - 5.00 mA |
Resolution: |
0.01 mA/step |
Accuracy: |
± (2% of setting + 0.02 mA) |
DC Current Display |
Range: |
0.00 - 5.00 mA |
Resolution: |
0.01 mA/step |
Accuracy: |
± (2% of reading + 0.02 mA) |
Failure Detector |
Audible and Visual |
Output Ripple |
< 5% (6KVDC/5 mA at Resistive Load) |
Discharge Time |
< 200 ms |
Maximum Capacitive Load in DC Mode |
1μF < 1KV , 0.75μF < 2KV, 0.5μF < 3KV, 0.08μF < 4KV, 0.04μF < 5KV, 0.01μF < 6KV |
AC Wave Form |
Sine wave Distortion : < 2% THD ; Crest Factor= 1.3 – 1.5 |
Output Frequency |
Range: |
60 or 50 Hz, User Selection |
Output Regulation |
± (1 % of output + 5 volts) from no load to full load |
Dwell Timer |
Range: |
0, 0.2 - 999.9 sec, (0 = Constant) |
Resolution: |
0.1 sec increments |
Accuracy: |
± (0.1% + 0.05 sec) |
Ramp Timer |
Range: |
0.1 - 999.9 sec |
Resolution: |
0.1 sec increments |
Accuracy |
± (0.1% + 0.05 sec) |
INSULATION RESISTANCE TEST MODE |
Output Voltage |
Range: |
100 - 1000 Volts DC |
Resolution: |
1 volt/step |
Accuracy: |
± (2% of reading + 5 volts) |
Voltage Display |
Range: |
0 - 1000 V |
Resolution: |
1 volt/step |
Accuracy: |
± (2% of reading + 2 counts) |
Resistance Display |
Range: |
1 - 1000 MW (4 Digit, Auto Ranging) |
Resolution: |
0.01/0.1 MW |
Accuracy: |
± (3% of reading + 2 counts) at test voltage > 500V |
± (7% of reading +2 counts) at test voltage < 500V |
HI-Limit |
Range: |
0, 1 - 1000 MW (0 = Off) |
LO-Limit |
Range: |
1 - 1000 MW |
Delay Timer |
Range: |
0, 0.5 - 999.9 sec (0 = Constant) |
Resolution: |
0.1 sec/step |
Accuracy: |
± (0.1% + 0.05 sec) |
GROUND BOND TEST MODE |
Output Voltage |
Range: |
AC 6 volts Max. |
Output Frequency |
Range: |
60 or 50 Hz, User Selection |
Accuracy: |
±0.1% |
Output Current |
Range: |
3.0 - 30.0 Amps AC |
Resolution: |
0.1 Amp/step |
Accuracy : |
± (2 % of setting + 0.1A) |
Current Display |
Range: |
0.0 - 30.0 Amps |
Resolution: |
0.1 Amp/step |
Accuracy: |
± (3 % of reading + 0.1A) |
Resistance HI-Limit |
Range: |
0 - 510 mW |
Resistance Display |
Range: |
0 – 510 mW for 0.0 – 10.0 Amps |
0 – 200 mW for 10.1 – 25.0 Amps |
0 – 150 mW for 25.1 – 30.0 Amps |
Resolution: |
1 mW/step |
Accuracy: |
± (2 % of Setting + 2 mW ) |
Dwell Timer |
Range: |
0, 0.5 - 999.9 sec (0 = Constant) |
Resolution: |
0.1 sec/step |
Accuracy: |
± (0.1% + 0.05 sec) |
Lead Resistance Offset |
Max. Offset Capability: |
100 mW |
Resolution: |
1 mW / step |
Accuracy: |
± (2 % of Setting + 2 mW ) |
GENERAL SPECIFICATIONS |
PLC Remote Control |
Input - Test, Reset, Interlock & Recall Memory 1 - 6 |
Output - Pass, Fail, Test-in-Process |
Memory |
Allows storage of up to 6 different test programs and 6 step/each memory and single step mode. |
Security |
Lockout capability to avoid unauthorized access to test set-up programs. |
Calibration |
Software and adjustments are made through front panel. |
Mechanical |
Bench or rack mount with tilt up front feet. |
Optional Arc Detection Hardware |
ACW and DCW; Software adjustable Arc Level 0-9 |
Dimension |
280(W) x 89(H) x 430(D) mm |
Weight |
33 lbs. (15 kg) |