SPECIFICATIONS |
MODEL |
7110 |
7112 |
7120 |
7122 |
AC WITHSTAND VOLTAGE |
Output Rating |
5KVAC/12mA |
|
Range |
Resolution |
Accuracy |
Output Voltage, KVAC |
0-5.00 |
0.01 |
±(2% of setting + 5V) |
Output Frequency |
50Hz/60Hz ±100ppm, User Selection |
Output Waveform |
Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 |
Output Regulation |
± (1% of output + 5V), From no load to full load |
SETTINGS |
Hi-Limit AC Current, mA |
0.10 - 12.00 |
0.01 |
±(2% of setting + 2counts) |
Lo-Limit AC Current, mA |
0.00 - 12.00 |
0.01 |
±(2% of setting + 2counts) |
Ramp Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
Dwell Time, second |
0, 0.2 - 999.9 (0=continuous) |
0.1 |
Arc Detection |
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) |
DC WITHSTAND VOLTAGE (7120,7122 ONLY) |
Output Rating |
6KVDC/5mA |
Output Voltage, KVDC |
0 - 6.00 |
0.01 |
±(2% of setting + 5V) |
Output Ripple |
<5%(6KV/5mA at Resistive Load) |
SETTINGS |
Hi-LimitDC Current, mA |
0.02 - 5.00 |
0.01 |
±(2% of setting + 2counts) |
Lo-LimitDC Current, mA |
0.00 - 5.00 |
0.01 |
±(2% of setting + 2counts) |
Ramp Time, second |
0.1 - 999.9 |
0.1 |
±(0.1% + 0.05sec) |
Dwell Time, second |
0, 0.2 - 999.9 (0=continuous) |
0.1 |
Arc Detection |
0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) |
Discharge Time |
< 200 msec |
INSULATION RESISTANCE (7112,7122 ONLY) |
Output Rating |
1KVDC/1000MΩ |
Output Voltage, KVDC |
0.10 - 1.00 |
0.01 |
±(2% of setting + 3V) |
SETTINGS |
Hi-Limit Resistance, MΩ |
0, 1 - 1000 (0=OFF) |
1 |
±(3% of setting+2counts)at > 500VDC |
Lo-Limit Resistance, MΩ |
1 - 1000 |
1 |
±(7% of setting +2counts)at <500VDC |
Delay Time, second |
0, 0.5 - 999.9 (0=continuous) |
0.1 |
±(0.1% + 0.05sec) |
MEASUREMENT |
AC/DC Voltage, KV |
0 - 6.00 |
0.01 |
±(1.5% of reading + 1count) |
DC Voltage, KV (IR only) |
0.10 - 1.00 |
0.01 |
±(1.5% of reading + 3V) |
AC Current, mA |
0.10 - 12.00 |
0.01 |
±(2% of reading + 2counts) |
DC Current, mA |
0.02 - 5.00 |
0.01 |
±(2% of reading + 2counts) |
Resistance, MΩ |
1 - 1000 (AutoRange) |
0.001 0.01 0.1 |
±(3% of reading+2counts)at > 500VDC ±(7% of reading +2counts)at <500VDC |
GENERAL |
Input Voltage AC |
115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A Fast |
PLC Remote Control |
Input : Test , Reset |
Output : Pass , Fail , Processing |
Memory |
5 Sets Store/Recall of each Memory, 4 Test Mode Selectable (W, I, W~I, I~W Only 7112, 7122) |
Display |
16×2 LCD with back light |
Key Lock |
To prevent unauthorized alteration of the test Parameters |
Calibration |
Build-in software and external calibrated meters |
Environment |
0-40℃, 20-80%RH |
Dimension/Net Weight |
280mm(W)×89mm(H)×370mm(D)/ 9Kg |
STANDARD ACCESSORIES |
Power Cord(10A) |
×1 |
Fuses |
×3 (Including a spare contained in the fuse holder) |
Hipot Test Lead ,1.5m(1101) |
×1 |
Hipot Return Lead ,1.8m(1102) |
×1 |
*product specifications are subject to change without notice. |